Serious
Adult Portable Bed Rails Recalled Due to Risk of Serious Injury or Death from Entrapment and Asphyxiation; Violate Mandatory Standard for Adult Portable Bed Rails; Imported by Hepo Care Medical Equipment Online
This recall involves Hepo-branded adult portable bed rails, model 110039.
The hazard
The recalled bed rails violate the mandatory standard for adult portable bed rails because users can become entrapped within the bed rail or between the bed rail and the side of the mattress, posing a serious entrapment hazard and risk of death by asphyxiation. The bed rails also do not meet structural stability or retention strap requirements, posing a fall hazard. The bed rails' push pins and push pin holes are incorrectly sized, posing a laceration hazard. In addition, the bed rails do not bear the required hazard warning labels.
Injuries and incidents
None reported
The remedy
Consumers should stop using the recalled adult portable bed rails immediately and contact Hepo Care Medical Equipment Online for a full refund. Consumers will be asked to destroy the bed rails by dismantling the bed rails, cutting off the storage bag and writing "RECALLED" on the upper and lower rails with permanent marker, take a photo of the destroyed rails and upload the photo to www.hepolife.com/bed-rail-safety-recall or email the photo to manager@lqxmedical.com.
Recall details
Have this product?
Follow the remedy in the official notice. CPSC lists the exact models and where it was sold.
Open CPSC noticeCompany pages: Mobility Source Medical Technology Co., Ltd
Source: CPSC. Defect Index is independent and not affiliated with CPSC.